Biomechanics breakthrough
 

IDEEA Accuracy

Capabilities and Accuracy

Accuracy of Speed Measurement

  The accuracy of speed measurement is 99.3±3.6% (N=112). Speed predicted by IDEEA compared with the actual speed while a subject (male, 72kg, 174cm, 43yr) walks and runs on a treadmill.  


Energy expenditure and work estimation

  The accuracy in energy expenditure estimation is 98.1±5.3%. Comparison of EE predicted by IDEEA and measured by a room calorimeter. Error = -2.52%, absolute error |error| = 6.9%. The major error was caused by low efficiency during continuous sit/stand cycles.  


Cumulative EE (total calories consumed)

  Cumulative results (integration) of the same measurement above. Note error is less than 3%, due to the cancellation of positive and negative errors.  


Accuracy in Activity Type Detection (76 subjects)

(Detailed info on accuracy validation can be found from Obesity Research Journal, Vol. 11, No. 1, pp 33-40, Jan. 2003)
Primary Secondary Accuracy Primary Secondary Accuracy

Sitting

Upright (normal) 99.40% Leaning Left shoulder against wall 98.71%
Left leg over right leg 99.11% Right shoulder against wall 100.00%
Right leg over left leg 98.55% Two elbows on a counter 99.28%
Elbows on knees 99.28% Average 99.33%
Left heel up 99.28%
Right heel up 100.00% Lying Down Facing up 100.00%
Both heels up 99.28% On right shoulder 99.12%
Both feet elevated 98.55% Facing down 99.28%
Average 99.18% On left shoulder 99.28%
Average 99.23%
Standing Upright (normal) 99.03%
Left foot on a step 100.00% Limb movement Sit move left leg 99.60%
Right foot on a step 99.57% Sit move right leg 100.00%
Average 99.53% Sit move both legs 99.26%
Stand move left leg 98.16%
Reclining Both feet on the ground 95.63% Stand move right leg 99.90%
Left leg over right leg 99.23% Average 99.14%
Right leg over left leg 96.84%
Average 97.23%

Accuracy in Gait Identification (69 subjects)

(Detailed info on accuracy validation can be found from Obesity Research Journal, Vol. 11, No. 1, pp 33-40, Jan. 2003)
Type of Gait Actual Number Number Detected Rate of Detection
Walking 16,179 16,131 99.66 %
Running 10,421 10,341 98.99 %
Climbing Stairs 3,168 3,119 98.36 %
Descending Stairs 3,168 3,120 98.14 %
Jumping 621 600 96.60 %
Average

98.35 ± 1.14 %

(mean ± SD)

Go back